MALDI-TOF-MS analysis of antibody-antigen complexes

Editor:
Prof. Dr. Christiane Zaborosch

Head of Center for Biochemistry
Lecturer in Biochemistry
christiane.zaborosch@zhaw.ch

 Introduction

In the development of new drugs, it is of interest in pharmaceutical research to determine the size of antibody-antigen complexes that have formed, since it is assumed that large complexes are cleared more quickly by the body, and thus the antigen is removed faster. Antibody-antigen complexes (mAb-Ag) can be analysed using MALDI-TOF-MS. However, non-covalent complexes are difficult to detect using MALDI-TOF-MS without sample preparation, as dissociation of the complex occurs during laser irradiation and co-crystallization with the matrix. In order to prevent dissociation during the MALDI process, chemical stabilization can be carried out by means of covalent cross-linking. For cross-linking of the antibodies with the antigen a homobifunctional, amine-reactive cross-linker BS(PEG)5 was employed (spacer: 21.7 Å). The crosslinker reacts with the free N-terminus and the ε-amino group of lysine side chains of the proteins. The nucleophilic N atom of the protein’s primary amine breaks down the carbonyl group of the cross-linker and an amide bond is formed by nucleophilic substitution. To analyse the major cross-linked protein complexes (> 150 kDa) with MALDI-TOF-MS, an HM1 high-mass ion conversion detector from CovalX was used.

Binding experiment with one mAb

To study the mAb-Ag complexes, a model system with IL-1β (Ag) and two anti-human IL-1β monoclonal antibodies XO01_γ2 and BS01 (mAb) was used. In the binding experiment XO01_γ2 or BS01 and IL-1β were mixed at the ratio of 1:2. The complexes formed from Ag and mAb were crosslinked through the addition of BS(PEG)5 in a 56-fold molar excess.

MALDI-Spectra
Figure 1: High-mass MALDI mass spectra of the binding experiment of IL-1β and the mAb XO01_γ2. Ratio 1:2 mAb:Ag. A: before addition of the crosslinkers (control); B: after addition of BS(PEG)5, masses after removal of the crosslinker. Matrix: sinapic acid; Detector: HM1

In the control spectrum (Fig. 1A) the peaks of the antigen IL-1β occurred at 17.4 kDa and those of the mAb XO01_γ2 at 146.3 kDa. The peaks at 163.3 kDa and 180.3 kDa corresponded to complexes without crosslinking. In the spectrum after crosslinking (Fig. 1B), two very high intensity peaks were observed at 163.4 kDa and 180.2 kDa, which could be attributed to the [XO01_γ2*1IL-1β] and [XO01_γ2*2IL-1β] complexes. The mass of the crosslinker was subtracted from the calculated masses of the complexes. BS01 was also crosslinked with IL-1β and analysed (data not shown).

Multi-binding experiment with two mAbs

In the multi-binding experiment larger mAb-Ag complexes were generated and the exact masses determined using MALDI-TOF MS. The two mAbs which are directed against different epitopes were mixed with IL-1β in a ratio of 1:2:1 (mAb1:Ag:mAb2) and then crosslinked with the crosslinker BS(PEG)5 in a 56-fold molar excess.

MALDI-Spectrum
Figure 2: High-mass MALDI mass spectrum of the multi-binding experiment of IL-1β and the mAb XO01_γ2 and BS01 after crosslinking with BS(PEG)5 at a ratio of 1:2:1 mAb1:Ag:mAb2. Masses shown after removal of the crosslinker. Matrix: sinapic acid; Detector: HM1

On the basis of the molecular masses, the peaks in the spectrum (Fig. 2) were attributed to the various mAb-Ag complexes (Tab. 1). Uncomplexed antigen was not detected; IL-1 β was therefore fully bound into complexes.

Table 1
Table 1: Allocation of the mAb-Ag complexes formed

Conclusion

Using the crosslinker BS(PEG)5 the two mAbs XO01_γ2 and BS01 were covalently crosslinked with the antigen IL-1β and the resulting mAb-Ag-complexes were detected using MALDI-TOF-MS.

[ratings]

Subscribe
Notify of
guest
8 Comments
oldest
newest most voted
Inline Feedbacks
View all comments
Roland Josuran
Editor
9. December 2014 21:49

Thank you, Miriam, for your contribution. Have you ever tried to calculate the theoretical molecular mass of the antibody-antigen complexes? There’s an opportunity to consider multiple amino acid chains and custom defined modifications with Prot pi | Protein Tool.

Amit
Amit
24. June 2015 8:25

Hi How to rule out Gas phase dimers and oligomers in MALDI MS spectra??

pk service
1. September 2015 9:13

Hi, is it possible to detect non-covalent complexes using MALDI-TOF-MS?

Elsa
Elsa
24. June 2019 18:01

Hi Miriam, really nice work! Could you please share some of your sample prep methods? I saw that you used Sinapinic acid as your matrix. What was the matrix reconstituted with and did you premix the matrix with your sample prior to deposition on the plate?